Z-2224-2023 Class II Ongoing

Recalled by Datascope Corp. — Fairfield, NJ

FDA device recall Z-2224-2023 was initiated by Datascope Corp. on June 23, 2023 and is designated Class II. Reason for recall: During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral arter… The recall status is ongoing. Affected quantity: 321,609 total kits.

Recall Details

Product Type
Devices
Report Date
August 2, 2023
Initiation Date
June 23, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
321,609 total kits

Product Description

REINFORCED INTRODUCER SET, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. D684-00-0403-05 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. D684-00-0403-06 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. D684-00-0403-10

Reason for Recall

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Distribution Pattern

Domestic distribution nationwide. International distribution worldwide.

Code Information

Product Description UDI REINFORCED INTRODUCER SET FOR LINEAR 7.5Fr. & MEGA 7.5Fr. 10607567106656 REINFORCED INTRODUCER SET FOR SENSATION 7Fr. 10607567106694 REINFORCED INTRODUCER SET FOR SENSATION PLUS & MEGA 8Fr. 10607567107943