Z-2250-2012 Class I Terminated

Recalled by Newport Medical Instruments Inc — Costa Mesa, CA

FDA device recall Z-2250-2012 was initiated by Newport Medical Instruments Inc on Aug 3, 2012 and is designated Class I. Reason for recall: May emit a continuous high priority alarm and the ventilator may stop ventilating, due to a component failure on the control board. The recall status is terminated (terminated Apr 14, 2014). Affected quantity: 72.

Recall details

Product Type
Devices
Report Date
Sep 12, 2012
Initiation Date
Aug 3, 2012
Termination Date
Apr 14, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
72

Product description

Newport HT70 and HT70 Plus Ventilators, Model Number: HT70 and HT70 Plus. Product Usage: The HT70 ventilator system is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. Specifically, the HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs.). The HT70 is a restricted medical device intended for use by qualified, trained personnel under the direction of a physician. The HT70 is suitable for use in hospital, sub-acute, emergency room, and home care environments, as well as for transport and emergency response applications.

Reason for recall

May emit a continuous high priority alarm and the ventilator may stop ventilating, due to a component failure on the control board.

Distribution pattern

Worldwide distribution: USA (nationwide) and countries of: Brazil, Chile, Egypt, France, Hong Kong, India, Japan, Norway, Philippines and South Africa.

Code information

Model Number: HT70 and HT70 Plus. Serial numbers: N12HT700514226 to N12HT700514231, N12HT700514233 to N12HT700514236, N12HT700514238 to N12HT700614280, N12HT700614282 to N12HT700614299, N12HT700614301 to N12HT700614309, N12HT720410266 to N12HT720410268, N12HT720410271, N12HT720410273, N12HT720410282, N12HT720410283, N12HT720410286, N12HT720410288, N12HT720410289, N12HT720410290, N12HT720410294 to N12HT720510300, N12HT720510302, N12HT720510322 to N12HT720510330, N12HT720510332 to N12HT720510344,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2250-2012
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Newport Medical Instruments Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Newport Medical Instruments Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Newport Medical Instruments Inc →

Source: openFDA record for Z-2250-2012. Report a data issue: hello@regdatalab.com.