Z-2250-2017 Class II Terminated

Recalled by Circadiance LLC — Export, PA

FDA device recall Z-2250-2017 was initiated by Circadiance LLC on May 1, 2017 and is designated Class II. Reason for recall: Circadiance has determined that it is possible for certain Smart Monitor 2PS/PSL monitors to exhibit intermittent operation of the nurse call feature. The firm has updated the device to increase the… The recall status is terminated (terminated Feb 5, 2018). Affected quantity: 1,242 units.

Recall details

Product Type
Devices
Report Date
Jun 14, 2017
Initiation Date
May 1, 2017
Termination Date
Feb 5, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,242 units

Product description

SmartMonitor 2 PS/PSL, Monitor, Apnea, Facility USE. Intended for use in the continuous monitoring of respiration, heart rate, and SP02 of infant, pediatric and adult patients.

Reason for recall

Circadiance has determined that it is possible for certain Smart Monitor 2PS/PSL monitors to exhibit intermittent operation of the nurse call feature. The firm has updated the device to increase the "Nurse Call" feature to ensure monitor alarms are transmitted to compatible nurse call systems.

Distribution pattern

Nationwide Distribution.

Code information

Unique Device Identifier (UDI) 1030271, 1038140, 1038141, U1030271, and U1038140. All serial numbers of units with "REF", "MODEL" or standalone part number 1030271, 1038140, 1038141, U1030271 or U1038140 located on the products label are impacted by this notice. Monitors with the above part numbers that have a serial number starting with "CP" are NOT affected by this issue.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2250-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Circadiance LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2250-2017. Report a data issue: hello@regdatalab.com.