Z-2250-2017 Class II Terminated

Recalled by Circadiance LLC — Export, PA

Recall Details

Product Type
Devices
Report Date
June 14, 2017
Initiation Date
May 1, 2017
Termination Date
February 5, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,242 units

Product Description

SmartMonitor 2 PS/PSL, Monitor, Apnea, Facility USE. Intended for use in the continuous monitoring of respiration, heart rate, and SP02 of infant, pediatric and adult patients.

Reason for Recall

Circadiance has determined that it is possible for certain Smart Monitor 2PS/PSL monitors to exhibit intermittent operation of the nurse call feature. The firm has updated the device to increase the "Nurse Call" feature to ensure monitor alarms are transmitted to compatible nurse call systems.

Distribution Pattern

Nationwide Distribution.

Code Information

Unique Device Identifier (UDI) 1030271, 1038140, 1038141, U1030271, and U1038140. All serial numbers of units with "REF", "MODEL" or standalone part number 1030271, 1038140, 1038141, U1030271 or U1038140 located on the products label are impacted by this notice. Monitors with the above part numbers that have a serial number starting with "CP" are NOT affected by this issue.