Z-2262-2020 Class II
Recall Details
- Product Type
- Devices
- Report Date
- June 17, 2020
- Initiation Date
- October 4, 2019
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 461
Product Description
Model Number: 06917259001 UDI: 04015630976010 - Product Usage: The System s intended use environment includes anatomic pathology (AP) laboratories and histology facilities within hospital, clinic, reference, or private lab settings. The System is to be operated by qualified users in
Reason for Recall
The firm became aware of a cleaning solution leaking issue from the middle staining module of the Ventana HE 600 instrument, used for histologic section samples in anatomic pathology laboratory settings, which caused fluid to travel down to the electrical wiring connector of the Coverslipper solvent shield circuit board. This can cause an electrical short and result in burned components.
Distribution Pattern
Worldwide distribution - US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, MD, MI, MA, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Costa Rica, Croatia, Denmark, Ecuador, El Salvador, Finland, France, Germany, Hong Kong, Ireland, Italy, Japan, Kuwait, Lebanon, Malaysia, Norway, Peru, Poland, Saudi Arabia, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam.
Code Information
All lots