Z-2265-2025 Class III Ongoing

Recalled by DRG International, Inc. — Springfield, NJ

FDA device recall Z-2265-2025 was initiated by DRG International, Inc. on Jun 4, 2025 and is designated Class III. Reason for recall: An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months… The recall status is ongoing. Affected quantity: 47 units.

Recall details

Product Type
Devices
Report Date
Aug 13, 2025
Initiation Date
Jun 4, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
47 units

Product description

Brand Name: T4 Total ELISA Product Name: T4 Total ELISA Model/Catalog Number: EIA-4568 Product Description: T4 Total ELISA

Reason for recall

An incoming complaint reported that the expiration date on the kit box label and Certificate of Analysis (CoA) exceeded the shelf life of the standards and control solutions in the kit by 2 months (i.e., 2026-03-31 instead of 2026-01-31). The expiration dates on the bottle labels were correct (i.e., 2026-01-31).

Distribution pattern

Worldwide - US Nationwide distribution in the states of CA, MA, PA and the countries of Germany, Hong Kong, Moldova.

Code information

Model/Catalog Number: EIA-4568; UDI-DI: 04048474045680; Lot number: 304K034;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2265-2025
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2265-2025. Report a data issue: hello@regdatalab.com.