Z-2280-2012 Class II Terminated

Recalled by Advanced Neuromodulation Systems Inc. — Plano, TX

Recall Details

Product Type
Devices
Report Date
September 5, 2012
Initiation Date
December 19, 2011
Termination Date
July 24, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
26,388 units

Product Description

Eon (Model 3716) IPG is a dual channel; 8-contact receiver designed to be connected to 4, 8, or 16 electrode leads or a lead extension. Product Usage: The Eon and Eon Mini Neurostimulation Systems are indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.

Reason for Recall

As of 11/30/2011, St. Jude Medical has received 110 patient complaints of warmth or heating at the implantable pulse generators (IPGs) implant site during charging for the Eon IPG and 116 reports of similar symptoms for the Eon Mini IPG, respectively.

Distribution Pattern

Nationwide within the US, and Internationally to Argentina, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Isreal, Italy, Japan, Mexico, Netherlands, Puerto Rico, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

Code Information

All lots of the Eon (Product Code 65-3716)