Z-2304-2012 Class II Terminated

Recalled by Sendx Medical Inc — Carlsbad, CA

Recall Details

Product Type
Devices
Report Date
September 5, 2012
Initiation Date
June 6, 2012
Termination Date
September 4, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
ABL80 FLEX is 1680, ABL80 CO-OX is 1759

Product Description

ABL80 FLEX (with FLEX software) and ABL80 FLEX CO-OX (with CO-OX software) - pH Blood Gas and Electrolytes Analyzer System. Intended Use The ABL80 FLEX (with FLEX software) and ABL80 FLEX CO-OX (with CO-OX software) are portable, automated analyzers that measure pH, blood gases, electrolytes, glucose and oximetry (ABL80 FLEX CO-OX only) in whole blood. The ABL80 FLEX (with FLEX software) and ABL80 FLEX CO-OX (with CO-OX software) analyzer systems are intended for use by trained technologists, nurses, physicians and therapists. They are intended for use in a laboratory environment, near patient or point-of-care setting.

Reason for Recall

The firm recalled because the glucose measurements from a patient sample that the customer felt was too low and did not reflect the patient's clinical condition.

Distribution Pattern

Worldwide Distribution - US Nationwide including the state of OH.

Code Information

Model Numbers: 393-839 (with 933-082 software) and 393-841 (with 933-132 software) All Serial Numbers