Z-2304-2012 Class II Terminated

Recalled by Sendx Medical Inc — Carlsbad, CA

FDA device recall Z-2304-2012 was initiated by Sendx Medical Inc on Jun 6, 2012 and is designated Class II. Reason for recall: The firm recalled because the glucose measurements from a patient sample that the customer felt was too low and did not reflect the patient's clinical condition. The recall status is terminated (terminated Sep 4, 2012). Affected quantity: ABL80 FLEX is 1680, ABL80 CO-OX is 1759.

Recall details

Product Type
Devices
Report Date
Sep 5, 2012
Initiation Date
Jun 6, 2012
Termination Date
Sep 4, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
ABL80 FLEX is 1680, ABL80 CO-OX is 1759

Product description

ABL80 FLEX (with FLEX software) and ABL80 FLEX CO-OX (with CO-OX software) - pH Blood Gas and Electrolytes Analyzer System. Intended Use The ABL80 FLEX (with FLEX software) and ABL80 FLEX CO-OX (with CO-OX software) are portable, automated analyzers that measure pH, blood gases, electrolytes, glucose and oximetry (ABL80 FLEX CO-OX only) in whole blood. The ABL80 FLEX (with FLEX software) and ABL80 FLEX CO-OX (with CO-OX software) analyzer systems are intended for use by trained technologists, nurses, physicians and therapists. They are intended for use in a laboratory environment, near patient or point-of-care setting.

Reason for recall

The firm recalled because the glucose measurements from a patient sample that the customer felt was too low and did not reflect the patient's clinical condition.

Distribution pattern

Worldwide Distribution - US Nationwide including the state of OH.

Code information

Model Numbers: 393-839 (with 933-082 software) and 393-841 (with 933-132 software) All Serial Numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2304-2012
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Sendx Medical Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2304-2012. Report a data issue: hello@regdatalab.com.