Z-2342-2025 Class II Ongoing
FDA device recall Z-2342-2025 was initiated by GE Medical Systems, LLC on Jul 2, 2025 and is designated Class II. Reason for recall: GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape… The recall status is ongoing.
Recall details
- Product Type
- Devices
- Report Date
- Sep 3, 2025
- Initiation Date
- Jul 2, 2025
- Voluntary/Mandated
- FDA Mandated
Product description
GE HealthCare OEC 9900 Elite C-arm Systems.
Reason for recall
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
Distribution pattern
US Nationwide distribution.
Code information
GTIN: 00840682114349
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2342-2025
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026