Z-2360-2023 Class II Ongoing
FDA device recall Z-2360-2023 was initiated by Siemens Medical Solutions USA, Inc on June 19, 2023 and is designated Class II. Reason for recall: Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may b… The recall status is ongoing. Affected quantity: 7 units.
Recall Details
- Product Type
- Devices
- Report Date
- August 16, 2023
- Initiation Date
- June 19, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 units
Product Description
Reason for Recall
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Italy, Japan, Latvia, Luxembourg, Macedonia, Mexico, Monaco, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.