Z-2360-2023 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

Recall Details

Product Type
Devices
Report Date
August 16, 2023
Initiation Date
June 19, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7 units

Product Description

ARTIS icono floor-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327700

Reason for Recall

Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Belgium, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Italy, Japan, Latvia, Luxembourg, Macedonia, Mexico, Monaco, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.

Code Information

UDI-DI: 4056869149325 Serial Numbers: 170561 170549 170559 170546 170552 170571 170570