Z-2373-2018 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
July 11, 2018
Initiation Date
May 30, 2018
Termination Date
May 28, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
101

Product Description

Zimmer¿ M/L Taper Hip Prosthesis With Kinective¿ Technology Cementless, Size 22.5; 00-7713-022-00

Reason for Recall

M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris.

Distribution Pattern

Products were distributed solely to Japan.

Code Information

00111778 60940196 61025425 61063904 61097911 61120383 61134445 61619526 62420837 62798037