Z-2391-2020 Class II Terminated

Recalled by Medtronic Sofamor Danek USA Inc — Memphis, TN

Recall Details

Product Type
Devices
Report Date
June 24, 2020
Initiation Date
April 24, 2020
Termination Date
March 31, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product Description

Medtronic Bone Void Filler, Large - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit contains the necessary components for mixing of the bone void filler.

Reason for Recall

Product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may increase the required surgical and anesthesia time for the patient. If the device seeps or creeps from its intended position prior to setting and creates incomplete void or gap filling, possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.

Distribution Pattern

Nationwide distribution to CA, FL, IL, IN, KY, MN, MO, MT, OH, PA, TX, VA, WA.

Code Information

Product number: C07B; GTIN: 00813845020511; Lot Number: 1127302