Z-2399-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 24, 2020
- Initiation Date
- May 20, 2020
- Termination Date
- May 13, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- Domestic distribution 65 devices. International distribution 12 devices. Total of devices distributed 77.
Product Description
Leica ARveo and M530 OH6 microscope systems.
Reason for Recall
Possible unintended interruption of surgical procedures due to equipment shutdown, related to changes in voltage.
Distribution Pattern
Domestic distribution: AL,AR,AZ, CA, CT, DC, DE, FL, IL, KY, LA, MI, MN, MO, NC, NY, OH. OR. TN, TX, UT, VA, WA, WI ,WV. International distribution: Japan
Code Information
M530 OH6 Serial Numbers for Model # 10448680 OH6-016215, OH6-042317, OH6-032016, OH6-009815,OH6-044518 and OH6-030016. Arveo Serial Numbers for Model # 10448999 OHA-006318, OHA-010719, OHA-011919, OHA-000818, OHA-009119, OHA-000718, OHA-014919, OHA-016619, OHA-015919, OHA-007818, OHA-013519, OHA-001218, OHA-012219, OHA-017319, OHA-000418, OHA-017119, OHA-013719, OHA-009919, OHA-017219, OHA-017419, OHA-002218, OHA-000318, OHA-014819, OHA-015219, OHA-003718, OHA-001018, OHA-016019, OHA-012919, OHA-012619, OHA-011719, OHA-012119, OHA-010819, OHA-016119, OHA-009219, OHA-002618, OHA-015719, OHA-012019, OHA-009519, OHA-000618, OHA-008518, OHA-004818, OHA-012719, OHA-010619, OHA-001118, OHA-010419, OHA-016319, OHA-016519, OHA-010219, OHA-012519, OHA-015419, OHA-009018, OHA-016419, OHA-016719, OHA-008918, OHA-007618, OHA-009619, OHA-012319, OHA-010519, OHA-005418.