Z-2465-2023 Class II Ongoing

Recalled by Cardiac Assist, Inc — Pittsburgh, PA

Recall Details

Product Type
Devices
Report Date
September 6, 2023
Initiation Date
June 30, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part number: 5800-1000. Note: The 5800-1000 model number is an individually packaged component contained within the Kit numbers: 5830-2916 LS TANDEMLIFE KIT DL29 5840-2417 LS TANDEMLIFE KIT - V24, A17

Reason for Recall

Epoxy used during manufacturing of the LifeSPARC Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,

Distribution Pattern

US Nationwide distribution in the states of IL, FL.

Code Information

S/N and UDI: (1)Kit 5830-2916 Pump SN 00876971 UDI: ( 01)00814112020609(11)230411(17)250219(21)00876971 (2)Kit 5840-2417 Pump SN 00876972 UDI: (01)00814112020609(11)230411(17)250219(21)00876972