Z-2478-2025 Class II Ongoing

Recalled by WASATCH PHOTONICS — Morrisville, NC

FDA device recall Z-2478-2025 was initiated by WASATCH PHOTONICS on August 4, 2025 and is designated Class II. Reason for recall: Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these … The recall status is ongoing. Affected quantity: 192.

Recall Details

Product Type
Devices
Report Date
September 17, 2025
Initiation Date
August 4, 2025
Termination Date
N/A
Voluntary/Mandated
FDA Mandated
Product Quantity
192

Product Description

Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

Reason for Recall

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

Distribution Pattern

U.S.

Code Information

all codes included in the model numbers.