Z-2507-2018 Class II Ongoing

Recalled by COVIDIEN MEDTRONIC — NORTH HAVEN, CT

FDA device recall Z-2507-2018 was initiated by COVIDIEN MEDTRONIC on May 22, 2018 and is designated Class II. Reason for recall: Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Jul 25, 2018
Initiation Date
May 22, 2018
Voluntary/Mandated
Voluntary: Firm initiated

Product description

Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT

Reason for recall

Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

Distribution pattern

Nationwide U.S.A. Internationally to: Albania, Algeria, Austria, Belgium, Bulgaria, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Iceland, Ireland, Israel, Italy, Ivory Coast, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Malta, Martinique, Moldova, Morocco, Namibia, Netherlands, New Caledonia, Norway, Pakistan, Poland, Portugal, Reunion, Romania, Russian federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, Turkey, United Arab Emirates & United Kingdom

Code information

N7L0852X, N7M0130X, N8A0145X, N8A0336X, N7L0931X, N7M0639X, N8A0796X & N8B0993X

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2507-2018
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for COVIDIEN MEDTRONIC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2507-2018. Report a data issue: hello@regdatalab.com.