Z-2519-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- September 10, 2025
- Initiation Date
- July 31, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 200 units (20 packs containing 10 units per pack)
Product Description
Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Model/Catalog Number: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Software Version: Not Applicable Product Description: Sterile Cover Component: Lumicell Direct Visualization System (DVS) / PMA # P230014
Reason for Recall
The potential of the elastic bands included in the kit detaching or fracturing during surgical procedures. As a result, these failures may pose a risk of retained foreign bodies in the surgical cavity and potentially leading to surgical site infections.
Distribution Pattern
US Nationwide distribution in the states of CA, FL, MA, NC, TX.
Code Information
Model/Catalog Number: 900-00218 UDI-DIs: Sterile Cover 10-Pack Shipper Box Assembly: (01)0 0860008 53781 2, 840-00204 Assy, Dispenser Carton Sterile Cover: (01)0 0860003 25251 2, 810-00210 Assy, Outer Tray Sterile Cover: (01) 0 0860003 2520 5, Lot numbers: 43930, 43931, 43932, 43933, 42756, 44349, 44349-01, 44349-02, 44349-03, 44349-04;