Z-2522-2015 Class II Terminated

Recalled by Halyard Health, Inc — Alpharetta, GA

Recall Details

Product Type
Devices
Report Date
September 9, 2015
Initiation Date
July 29, 2015
Termination Date
January 12, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

KimVent* Closed Suction System for Neonates/Pediatrics, 6 F, Elbow (Product Code 206); KimVent* Closed Suction System for Neonates/Pediatrics, 8 F, Elbow (Product Code 208); KimVent* Closed Suction System for Pediatrics, 10 F, Elbow (Product Code 210); KimVent* Closed Suction System for Neonates/Pediatrics, 10 F, Elbow (Product Code 2103); KimVent* Closed Suction Systems for Neonates/Pediatrics, 8 F, Elbow (Product Code 8313); KimVent* Closed Suction System for Neonates/Pediatrics, 8 F, Elbow (Product Code 20083) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.

Reason for Recall

The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.

Distribution Pattern

Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada

Code Information

Product Code 206 - M5076T503, M5082T501, M5091T504; Product Code 208 - M5089T506, M5096T507, M5103T509; Product Code 210 - M5091T509; Product Code 2103 - M5082T401, M5089T401, M5138T402; Product Code 8313 - M5089T508; Product Code 20083 - M5082T403, M5082T404, M5089T402, M5103T401, M5166T402