Z-2522-2023 Class II Ongoing

Recalled by Aomori Olympus Co., Ltd. — Kuroishi, N/A

Recall Details

Product Type
Devices
Report Date
September 13, 2023
Initiation Date
July 27, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,048 units

Product Description

Olympus Distal Attachment (in Olympus Disposable EMR Kit)- Designed to be attached to the distal end of the endoscope. Disposable EMR Kit: K-001, K-002, K-003, K-004, K-011, K-012

Reason for Recall

The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002

Distribution Pattern

US Nationwide distribution.

Code Information

(1)K-001: Lot Numbers: 5835330 0XI, 0YI, 12I, 14I, 15I, 16I, 17I, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK, 31K, 32K. UDI: 4953170048562 (2) K-002 Lot Numbers: 5835430 0XI, 0YI, 12I, 13I, 14I, 16I, 17I, 27K, 2XK, 2YK, 2ZK 4953170048579 K-003 5835530 0XI, 0YI, 12I, 13I, 14I, 15I, 16I, 17I, 25K, 26K, 27K, 28K, 29K, 2XK, 2YK, 2ZK UDI:4953170048586 (3)K-004 Lot Numbers: 5835630 16I, 17I, 26K UDI: 4953170048593 (4)K-011 Lot Numbers: N2485230 0XI, 15I, 17I UDI: 4953170214035 (5) K-012 Lot Numbers: N2485330 16I, 17I UDI: 4953170214042