Z-2530-2016 Class II Terminated

Recalled by Danville Materials, Inc. — San Ramon, CA

Recall Details

Product Type
Devices
Report Date
August 24, 2016
Initiation Date
June 24, 2016
Termination Date
December 9, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
~ 2123 units

Product Description

Perioscopy Sheath: Part number: 1008; The Perioscopy Sheath is a single-use sterile component used with the Perioscopy System, which is a medical device that features miniaturized digital video technology, and enables the healthcare professional to see magnified details of the tooth anatomy and deposits below the gum line (i.e. calculus root fractures, restoration margins, tooth decay).

Reason for Recall

The Perioscopy Sheaths are currently provided with a label that indicates the Sheath has a 3 year shelf-life/expiry date. Danville Materials has become aware that validation data are not available to support the 3-year shelf-life/expiration dating claim.

Distribution Pattern

US and New Zealand, Australia, Canada, Japan, Israel, United Kingdom, Italy, Mexico.

Code Information

Part number 1008: Lot numbers: 1133; 1135; 1136; 1137; 1138; 1163; 1171; 1172; 1173; 1174; 1175; 1218; 1240; 1243; 1258; 1264; 1264-1; 1275; 1276; 1277; 1311; 1314; 1315; 1325; 1326.