Z-2537-2020 Class II Terminated

Recalled by Vitalconnect Inc. — San Jose, CA

Recall Details

Product Type
Devices
Report Date
July 15, 2020
Initiation Date
May 21, 2020
Termination Date
May 13, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7

Product Description

VistaSolution, physiological signal transmitter. A software graphical user interface for use by healthcare professionals to display physiological data collected by the wireless remote monitoring system in healthcare settings. This is a secondary, adjunct patient monitor and is not intended to replace existing standard-of-care patient monitoring.

Reason for Recall

A healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but no notification was found. A software defect was encountered causing the lack of notification. The defect could result in exposing a patient to potential hypoxia for a short window of time. The firm is correcting the software to bring the performance back into compliance with specifications.

Distribution Pattern

U.S. Nationwide distribution including in the states of NJ, NY, VA, OUS: None

Code Information

Catalog number: 6200, Vista Solution version 2.1, UDI: 00850490007139