Z-2557-2020 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 22, 2020
- Initiation Date
- June 16, 2020
- Termination Date
- May 9, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 345244
Product Description
BodyGuard Microset, REF: A120-161XPS
Reason for Recall
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
Distribution Pattern
U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT. OUS (Foreign): Canada, UK
Code Information
All lots within expiry date