Z-2588-2023 Class II Ongoing

Recalled by BioFire Diagnostics, LLC — Salt Lake City, UT

FDA device recall Z-2588-2023 was initiated by BioFire Diagnostics, LLC on Apr 25, 2023 and is designated Class II. Reason for recall: The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE¿ FILMARRAY¿ Gastrointestinal (GI) Panel The recall status is ongoing. Affected quantity: 61,069.

Recall details

Product Type
Devices
Report Date
Sep 20, 2023
Initiation Date
Apr 25, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
61,069

Product description

FILMARRAY Gastrointestinal (GI) Panel (Part No.: RFIT-ASY-0104 & RFIT-ASY-0116) Pouch Module Software

Reason for recall

The firm has identified an increased risk of obtaining false positive Cryptosporidium results from customers using the BIOFIRE¿ FILMARRAY¿ Gastrointestinal (GI) Panel

Distribution pattern

US:DE, AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: AD, AE, AI, AR, AT, AU, BD, BE, BF, BG, BH, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DK, DZ, EE, ES, FI, FR, GA, GB, GE, GH, GM, GR, GU, HK, HU, IC, ID, IE, IL, IN, IQ, IT, JP, KE, KR, KW, LV, MO, MX, MY, NL, NO, NP, NZ, PG, PH, PL, PT, QA, RS, SA, SE, SG, SI, TH, TN, TR, TW, VN, ZA, ZW

Code information

Model/UDI-DI: RFIT-ASY-0104/00815381020116, RFIT-ASY-0116/00815381020109 using software versions prior to FLM2-SFW-0040 v2.0.3.1/BFS0001-5539 v2.0.200.10

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2588-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for BioFire Diagnostics, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2588-2023. Report a data issue: hello@regdatalab.com.