Z-2613-2025 Class I Ongoing

Recalled by Medline Industries, LP — Northfield, IL

Recall Details

Product Type
Devices
Report Date
October 8, 2025
Initiation Date
July 31, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7913 eaches

Product Description

Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH

Reason for Recall

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Distribution Pattern

Domestic US distribution nationwide. International distribution pending. No international distribution.

Code Information

Item Number/UDI-DI D087031RH 10197344019557 LOTS EP240731 EP240814 EP240826 EP240903 EP240916 EP241003 EP241014 EP241028 EP241104 EP241112 EP241202 EP241209A EP241219 EP241224 EP250114 EP250116 EP250204 EP250219