Z-2644-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 5, 2017
- Initiation Date
- May 22, 2017
- Termination Date
- July 26, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 477
Product Description
Exactech 1.5" Novation Calcar Planer Guide Tip To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use. ech 1.5" Novation Calcar Planer Guide Tip
Reason for Recall
The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the potential for instrument breakage if bending or torsional forces are applied.
Distribution Pattern
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AL, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii., and to the countries of : Argentina, Australia, Austria, Brazil, China, Colombia, France, Germany, Guatemala, India, Italy, Japan, Spain, Switzerland, Turkey, and United Kingdom
Code Information
Catalog #: 161-30-00 with the manufacturing lots #'s 38878-001, 38878-002, 49777-001, 51825-005, 51825-006, 52224-007, 51825-006, 52224-005, 52224-006, 57167-001, 59775-003, 62580-004, 66464-003, 66464-004, 70720-001, 71530-001, 73477-002, 75153-001, and 74999-001.