Z-2644-2017 Class II Terminated

Recalled by Exactech, Inc. — Gainesville, FL

Recall Details

Product Type
Devices
Report Date
July 5, 2017
Initiation Date
May 22, 2017
Termination Date
July 26, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
477

Product Description

Exactech 1.5" Novation Calcar Planer Guide Tip To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use. ech 1.5" Novation Calcar Planer Guide Tip

Reason for Recall

The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the potential for instrument breakage if bending or torsional forces are applied.

Distribution Pattern

Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AL, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii., and to the countries of : Argentina, Australia, Austria, Brazil, China, Colombia, France, Germany, Guatemala, India, Italy, Japan, Spain, Switzerland, Turkey, and United Kingdom

Code Information

Catalog #: 161-30-00 with the manufacturing lots #'s 38878-001, 38878-002, 49777-001, 51825-005, 51825-006, 52224-007, 51825-006, 52224-005, 52224-006, 57167-001, 59775-003, 62580-004, 66464-003, 66464-004, 70720-001, 71530-001, 73477-002, 75153-001, and 74999-001.