Z-2665-2023 Class II Ongoing

Recalled by Qiagen GmbH — Hilden, N/A

FDA device recall Z-2665-2023 was initiated by Qiagen GmbH on Aug 22, 2023 and is designated Class II. Reason for recall: Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity… The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Oct 11, 2023
Initiation Date
Aug 22, 2023
Voluntary/Mandated
Voluntary: Firm initiated

Product description

therascreen ¿KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Rotor-Gene Q MDx instrument for the detection of seven somatic mutations in the human KRAS oncogene REF 870021 (US IVD)

Reason for recall

Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, FL, IL, IN, KS, MA, NC,NY,OR, PA, TN, TX and the countries of KR, CN

Code information

GTIN number: 04053228002048 All lots that have an expiry date of March 2023 or later

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2665-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Qiagen GmbH? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2665-2023. Report a data issue: hello@regdatalab.com.