Z-2666-2017 Class II Terminated
FDA device recall Z-2666-2017 was initiated by Exactech, Inc. on May 25, 2017 and is designated Class II. Reason for recall: Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reamin… The recall status is terminated (terminated July 26, 2019). Affected quantity: 2,069.
Recall Details
- Product Type
- Devices
- Report Date
- July 12, 2017
- Initiation Date
- May 25, 2017
- Termination Date
- July 26, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,069
Product Description
EXACTECH Equinoxe Modular Reverse Pilot-Tip Reamer, Catalog #: 321-25-38, Catalog #: 321-25-42 and Catalog #: 321-25-46
Reason for Recall
Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.
Distribution Pattern
AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, Austria, Brazil, Canada, China, France, Germany, Greece, India, Italy, Korea, Luxembourg, Singapore, South Africa, Spain, Sweden, Switzerland, The Netherlands and United Kingdom
Code Information
All Lots