Z-2668-2020 Class II Terminated

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

Recall Details

Product Type
Devices
Report Date
July 29, 2020
Initiation Date
June 19, 2020
Termination Date
July 29, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5,296,132 total

Product Description

SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G (UDI 00801741087639), PA-0033G (UDI 00801741096433)

Reason for Recall

Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

Distribution Pattern

Worldwide Distribution

Code Information

Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G (UDI 00801741087639), PA-0033G (UDI 00801741096433). Lot Numbers: REEN3159 REEP3067 REEQ4064 REDS3541 REDV3438 REEN2781 REEN2145 REEP3068 REDY0835 REDZ3103