Z-2677-2014 Class II Terminated

Recalled by Genesys Orthopedics Systems, LLC — Austin, TX

FDA device recall Z-2677-2014 was initiated by Genesys Orthopedics Systems, LLC on Apr 2, 2014 and is designated Class II. Reason for recall: The recall is being initiated because MK Precision (the Circular Lock Screw supplier) notified Genesys Spine that a portion (less than 20%) of the Circular Lock Screws in lot 11406 were manufactured… The recall status is terminated (terminated May 15, 2015). Affected quantity: 642.

Recall details

Product Type
Devices
Report Date
Oct 1, 2014
Initiation Date
Apr 2, 2014
Termination Date
May 15, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
642

Product description

Circular Lock Screws, part # G826, lot # 11406. The Lock Screws are part of the Genesys Spine TiLock Pedicle Screw System. The TiLock Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion).

Reason for recall

The recall is being initiated because MK Precision (the Circular Lock Screw supplier) notified Genesys Spine that a portion (less than 20%) of the Circular Lock Screws in lot 11406 were manufactured using Stainless Steel in the center Post instead of the specified titanium alloy.

Distribution pattern

Nationwide Distribution including NV, CA, NY, MO, TX, FL, NJ, GA, TN.

Code information

The devices being recalled have a catalog number of G826 and a lot number of 11406.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2677-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Genesys Orthopedics Systems, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Genesys Orthopedics Systems, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Genesys Orthopedics Systems, LLC →

Source: openFDA record for Z-2677-2014. Report a data issue: hello@regdatalab.com.