Z-2688-2016 Class II Terminated

Recalled by Invivo Corporation — Gainesville, FL

FDA device recall Z-2688-2016 was initiated by Invivo Corporation on May 6, 2016 and is designated Class II. Reason for recall: A diode was inadvertently not assembled into the system, which may result in coil overheating The recall status is terminated (terminated Aug 21, 2019). Affected quantity: 3.

Recall details

Product Type
Devices
Report Date
Sep 7, 2016
Initiation Date
May 6, 2016
Termination Date
Aug 21, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product description

Invivo Sentinelle Venguard Breast MRI Auxiliary Table/Tabletop with 8/16 Channel Coil Array for GE 1.5T/3T MRI Systems.

Reason for recall

A diode was inadvertently not assembled into the system, which may result in coil overheating

Distribution pattern

WI

Code information

Model #: 9896-032-14382; Serial #: 13 (production order 301128446), 14 (production order 301153586), and 15 (production order 301153587).