Z-2722-2020 Class I Terminated

Recalled by CareFusion 303, Inc. — San Diego, CA

FDA device recall Z-2722-2020 was initiated by CareFusion 303, Inc. on Jun 30, 2020 and is designated Class I. Reason for recall: Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with… The recall status is terminated (terminated Apr 22, 2024). Affected quantity: 7854.

Recall Details

Product Type
Devices
Report Date
Aug 12, 2020
Initiation Date
Jun 30, 2020
Termination Date
Apr 22, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7854

Product Description

Alaris SpO2 Module Model 8210 and Model 8220, a modular infusion pump and monitoring system

Reason for Recall

Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.

Distribution Pattern

Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH

Code Information

During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)