Z-2722-2020 Class I Terminated

Recalled by CareFusion 303, Inc. — San Diego, CA

Recall Details

Product Type
Devices
Report Date
August 12, 2020
Initiation Date
June 30, 2020
Termination Date
April 22, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7854

Product Description

Alaris SpO2 Module Model 8210 and Model 8220, a modular infusion pump and monitoring system

Reason for Recall

Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.

Distribution Pattern

Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH

Code Information

During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers.