Z-2754-2016 Class II Terminated

Recalled by Anthogyr — Sallanches

FDA device recall Z-2754-2016 was initiated by Anthogyr on Jun 20, 2016 and is designated Class II. Reason for recall: Following a FDA inspection showing that the firm failed to validate the design and process of its implants. The recall status is terminated (terminated Mar 28, 2017). Affected quantity: 10 units.

Recall details

Product Type
Devices
Report Date
Sep 14, 2016
Initiation Date
Jun 20, 2016
Termination Date
Mar 28, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10 units

Product description

Axiom 2.8 implants Sterile The Axiom REG implants are used as artificial root structures for replacement of missing teeth. They can be used for stabilization of removable prostheses or fixation of single tooth restorations or partial dentures. There are different diameters and different lengths for each diameter.

Reason for recall

Following a FDA inspection showing that the firm failed to validate the design and process of its implants.

Distribution pattern

US Distribution to sates of: NC, NY, and OR.

Code information

Code OP28100 Size 02.8 x 10.0 Code OP28120 Size 02.8 x 12.0 Code OP28140 Size 02.8 x 14.0

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2754-2016
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Anthogyr? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2754-2016. Report a data issue: hello@regdatalab.com.