Z-2760-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 9, 2017
- Initiation Date
- July 11, 2017
- Termination Date
- May 3, 2023
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 146 units
Product Description
Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence
Reason for Recall
Instruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to end test).
Distribution Pattern
Worldwide Distribution - USA (nationwide) Distribution
Code Information
Models : VCR and VMR.