Z-2762-2015 Class II Terminated

Recalled by Revmedx Inc — Wilsonville, OR

Recall Details

Product Type
Devices
Report Date
September 23, 2015
Initiation Date
August 12, 2015
Termination Date
September 17, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5004 units

Product Description

XGAUZE Trauma Dressing, catalog number FIN-0002-01. XGAUZE is a 3" x 36" z-folded rayon/polyester dressing embedded with individual compressed cellulose sponges. Upon contact with blood or exudate, the sponges expand, thus increasing the dressing's capacity to absorb fluids. XGAUZE is packed in a pre-formed sterile barrier system (pouch sealed on three sides); one side is a PET/foil/PE film, and the other side is a PET/aluminum oxide/PE film. Packages are vacuum sealed and terminally sterilized via gamma radiation.

Reason for Recall

The XGAUZE Trauma Dressing was recalled because the packages may acquire small tears or pinholes during shipping due to use of shipping containers that lacked adequate protective materials. If present, these small tears or pinholes may disrupt package integrity and render the device unsterile.

Distribution Pattern

US Distribution to the states of : Arizona, Idaho, North Carolina, Tennessee and Virginia.

Code Information

Lot ESO54391, Expiration Date 9/25/2020; Lot ESO54725, Expiration Date 3/18/2020; Lot ESO54726, Expiration Date 3/30/2020; Lot ESO54741, Expiration Date 4/6/2020; Lot ESO54742, Expiration Date 4/13/2020; Lot ESO54743, Expiration Date 4/16/2020; Lot ESO54744, Expiration Date 4/23/2020;