Z-2763-2017 Class II Terminated

Recalled by Fisher Diagnostics — Middletown, VA

FDA device recall Z-2763-2017 was initiated by Fisher Diagnostics on May 19, 2017 and is designated Class II. Reason for recall: An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs… The recall status is terminated (terminated Jun 29, 2018). Affected quantity: 27 units.

Recall details

Product Type
Devices
Report Date
Aug 2, 2017
Initiation Date
May 19, 2017
Termination Date
Jun 29, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27 units

Product description

Pacific Hemostasis FDP Latex Reagent, REF/Model Number 100651, UDI 00845275000870, 1 x 5 mL, IVD; --- Fisher Diagnostics, a division of Fisher Scientific Company, LLC, a part of Thermo Fisher Scientific Inc., Middletown, VA 22645-1905 USA

Reason for recall

An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs accurately at FDP concentrations greater than 20 ¿g/ml, there is an observed reduction in sensitivity when FDP levels are less than 20 ¿g/ml. This reduction in sensitivity has been tied to FDP Latex Reagent lot 890199 and may result in false negatives.

Distribution pattern

USA (nationwide) including Puerto Rico, and Internationally to Colombia

Code information

Lot Number 890199, Exp. 6/30/2018

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2763-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Fisher Diagnostics? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Fisher Diagnostics: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Fisher Diagnostics →

Source: openFDA record for Z-2763-2017. Report a data issue: hello@regdatalab.com.