Z-2763-2017 Class II Terminated

Recalled by Fisher Diagnostics — Middletown, VA

Recall Details

Product Type
Devices
Report Date
August 2, 2017
Initiation Date
May 19, 2017
Termination Date
June 29, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
27 units

Product Description

Pacific Hemostasis FDP Latex Reagent, REF/Model Number 100651, UDI 00845275000870, 1 x 5 mL, IVD; --- Fisher Diagnostics, a division of Fisher Scientific Company, LLC, a part of Thermo Fisher Scientific Inc., Middletown, VA 22645-1905 USA

Reason for Recall

An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs accurately at FDP concentrations greater than 20 ¿g/ml, there is an observed reduction in sensitivity when FDP levels are less than 20 ¿g/ml. This reduction in sensitivity has been tied to FDP Latex Reagent lot 890199 and may result in false negatives.

Distribution Pattern

USA (nationwide) including Puerto Rico, and Internationally to Colombia

Code Information

Lot Number 890199, Exp. 6/30/2018