Z-2765-2019 Class II Terminated
Recalled by Roche Diabetes Care, Inc.
— Indianapolis, IN
FDA device recall Z-2765-2019 was initiated by Roche Diabetes Care, Inc. on Jun 12, 2017 and is designated Class II. Reason for recall: Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on). The recall status is terminated (terminated Oct 30, 2020). Affected quantity: 26,600.
Recall details
- Product Type
- Devices
- Report Date
- Oct 9, 2019
- Initiation Date
- Jun 12, 2017
- Termination Date
- Oct 30, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 26,600
Product description
Accu-Chek Aviva II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
Reason for recall
Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
Distribution pattern
Worldwide distribution - US Nationwide distribution.
Code information
Serial numbers between xxx20000000 and xxx23284925, or less than xxx06335297 Updated serial numbers on 10/12/2019: Aviva II: Between XXX20000000 and XXX23284925 Updated 10/18/2019: Aviva II from XXX05000001 to XXX06335297 OR from XXX20000000 to XXX23284925
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2765-2019
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Roche Diabetes Care, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Roche Diabetes Care, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Roche Diabetes Care, Inc. →
Use it in ChatGPT
Get a free API key
How the connector works