Z-2767-2026 Class II Ongoing
FDA device recall Z-2767-2026 was initiated by PIE Medical Imaging B.V. on May 7, 2026 and is designated Class II. Reason for recall: User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the… The recall status is ongoing. Affected quantity: 165.
Recall details
- Product Type
- Devices
- Report Date
- Jul 29, 2026
- Initiation Date
- May 7, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 165
Product description
3mensio Workstation (Vascular Fenestrated) software
Reason for recall
User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information
Distribution pattern
Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.
Code information
Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2767-2026
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026