Z-2768-2020 Class II Ongoing

Recalled by Covidien Llc — Mansfield, MA

Recall Details

Product Type
Devices
Report Date
August 12, 2020
Initiation Date
June 26, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
375 units

Product Description

Beacon EUS Access System Preloaded Access 90degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.

Reason for Recall

This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.

Distribution Pattern

world wide distribution

Code Information

Model # DSA-090-01 UDI: 10884521722583 Lot #: F2510370X, F2510900X, F2512340X, F2512378X, F2512492X, F2512620X, F2513452X