Z-2773-2015 Class III Terminated

Recalled by Pointe Scientific, Inc. — Canton, MI

FDA device recall Z-2773-2015 was initiated by Pointe Scientific, Inc. on Dec 12, 2014 and is designated Class III. Reason for recall: Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported… The recall status is terminated (terminated Oct 29, 2015). Affected quantity: 93 kits.

Recall details

Product Type
Devices
Report Date
Sep 23, 2015
Initiation Date
Dec 12, 2014
Termination Date
Oct 29, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
93 kits

Product description

Pointe Scientific Creatinine Reagent Catalog number C7539-150 in vitro diagnostic C7539-150 kit includes one R1 and one R2 bottle. R1: HDPE plastic bottle. 120 ml fill. R2: HDPE plastic bottle. 90 ml fill.

Reason for recall

Creatinine kit C7539-150 contains Creatinine reagent R1 that is labeled as Creatinine reagent R2 (30ml) and reagent R2 vial is labeled as R1 (120ml). With manual procedure, no result will be reported as the control will be out of specifications.

Distribution pattern

Worldwide distribution. US nationwide, Singapore, Kenya, Bolivia, Mexico, Greece, Guyana, and Philippines.

Code information

C7539-150 Lot: 406201 Batch: 279 Expiration date: 2016-03

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2773-2015
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Pointe Scientific, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2773-2015. Report a data issue: hello@regdatalab.com.