Z-2777-2017 Class II Terminated
FDA device recall Z-2777-2017 was initiated by GE Inspection Technologies, LP on Jun 28, 2017 and is designated Class II. Reason for recall: GE Inspection Technologies, LP determined that the Blade|line cabinet x-ray system fails to comply with the cabinet x-ray performance standards emission limit of 0.5 milliRoentgen in one hour (21 CFR… The recall status is terminated (terminated Feb 6, 2020). Affected quantity: 13 units distributed in the US.
Recall details
- Product Type
- Devices
- Report Date
- Aug 9, 2017
- Initiation Date
- Jun 28, 2017
- Termination Date
- Feb 6, 2020
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 13 units distributed in the US
Product description
Reason for recall
Distribution pattern
US Distribution to the states of : AL, SC, OH, MI, PA, and Internationally to Germany and Singapore