Z-2777-2017 Class II Terminated

Recalled by GE Inspection Technologies, LP — Lewistown, PA

FDA device recall Z-2777-2017 was initiated by GE Inspection Technologies, LP on Jun 28, 2017 and is designated Class II. Reason for recall: GE Inspection Technologies, LP determined that the Blade|line cabinet x-ray system fails to comply with the cabinet x-ray performance standards emission limit of 0.5 milliRoentgen in one hour (21 CFR… The recall status is terminated (terminated Feb 6, 2020). Affected quantity: 13 units distributed in the US.

Recall details

Product Type
Devices
Report Date
Aug 9, 2017
Initiation Date
Jun 28, 2017
Termination Date
Feb 6, 2020
Voluntary/Mandated
FDA Mandated
Product Quantity
13 units distributed in the US

Product description

GE Inspection Technologies Blade|line Industrial Cabinet X-ray system

Reason for recall

GE Inspection Technologies, LP determined that the Blade|line cabinet x-ray system fails to comply with the cabinet x-ray performance standards emission limit of 0.5 milliRoentgen in one hour (21 CFR ¿ 1020.40(c)(1)(i)) underneath the bottom center of the system. The crawl space affected by the excess emission is approximately 22 inches high, normally covered by decorative panels, and is not intended to be accessed.

Distribution pattern

US Distribution to the states of : AL, SC, OH, MI, PA, and Internationally to Germany and Singapore

Code information

Serial No. TM 201317 TM 201312 TM 201323 TM 201331 TM 201401 TM 201329 TM 201404 TM 201413 TM 201414 TM 201415 TM 201424 TM 201425 158017R1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2777-2017
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for GE Inspection Technologies, LP? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for GE Inspection Technologies, LP: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about GE Inspection Technologies, LP →

Source: openFDA record for Z-2777-2017. Report a data issue: hello@regdatalab.com.