Z-2777-2017 Class II Terminated

Recalled by GE Inspection Technologies, LP — Lewistown, PA

Recall Details

Product Type
Devices
Report Date
August 9, 2017
Initiation Date
June 28, 2017
Termination Date
February 6, 2020
Voluntary/Mandated
FDA Mandated
Product Quantity
13 units distributed in the US

Product Description

GE Inspection Technologies Blade|line Industrial Cabinet X-ray system

Reason for Recall

GE Inspection Technologies, LP determined that the Blade|line cabinet x-ray system fails to comply with the cabinet x-ray performance standards emission limit of 0.5 milliRoentgen in one hour (21 CFR ¿ 1020.40(c)(1)(i)) underneath the bottom center of the system. The crawl space affected by the excess emission is approximately 22 inches high, normally covered by decorative panels, and is not intended to be accessed.

Distribution Pattern

US Distribution to the states of : AL, SC, OH, MI, PA, and Internationally to Germany and Singapore

Code Information

Serial No. TM 201317 TM 201312 TM 201323 TM 201331 TM 201401 TM 201329 TM 201404 TM 201413 TM 201414 TM 201415 TM 201424 TM 201425 158017R1