Z-2783-2020 Class II Terminated

Recalled by Flower Orthopedics Corporation — Horsham, PA

Recall Details

Product Type
Devices
Report Date
August 19, 2020
Initiation Date
July 8, 2020
Termination Date
October 8, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
25

Product Description

Calcaneus Plate, Extensile, Medium, Left, Catalog Number CPF 102 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.

Reason for Recall

The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).

Distribution Pattern

US Nationwide distribution including in the states of AZ, FL, and OH.

Code Information

Lot Number 1909660029, 1909660031; UDI 00840118115650