Z-2811-2026 Class II Ongoing

Recalled by Medtronic Perfusion Systems — Brooklyn Park, MN

FDA device recall Z-2811-2026 was initiated by Medtronic Perfusion Systems on June 24, 2026 and is designated Class II. Reason for recall: Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, isc… The recall status is ongoing. Affected quantity: 1,424 units.

Recall Details

Product Type
Devices
Report Date
August 5, 2026
Initiation Date
June 24, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,424 units

Product Description

EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322

Reason for Recall

Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

Distribution Pattern

Worldwide - US Nationwide distribution.

Code Information

UDI-DI: 20199150023251, Lot Numbers: 0232709184 0232745036 0232798416 0232862784 0232862787; UDI-DI: 00199150023257, Lot Numbers: 0232862787