Z-2824-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- August 22, 2018
- Initiation Date
- July 9, 2018
- Termination Date
- September 6, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 16
Product Description
Sorin Platinium CRT-D 1711 DF-1, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator
Reason for Recall
Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy.
Distribution Pattern
US distribution to: NJ, AZ, MS, OK, LA, PA, FL, MI, IL, and MS
Code Information
Model TDF033U, UDI GTIN 08031527015538 Serial Numbers: 619DK0FB, 618DK003, 619DK0C0, 622DK03B, 622DK12A, 624DK070, 624DK12F, 619DK021, 627DK0B0, 618DK0A4, 619DK097, 620DK065, 619DK09B, 619DK05F, 619DK088, 619DK0CC