Z-2889-2026 Class II Ongoing

Recalled by Beckman Coulter Inc. — Brea, CA

FDA device recall Z-2889-2026 was initiated by Beckman Coulter Inc. on Jun 26, 2026 and is designated Class II. Reason for recall: Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications." The recall status is ongoing.

Recall details

Product Type
Devices
Report Date
Aug 12, 2026
Initiation Date
Jun 26, 2026
Voluntary/Mandated
Voluntary: Firm initiated

Product description

REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument

Reason for recall

Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."

Distribution pattern

US Nationwide distribution including in the states of AK, AL, AR, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WI, WV, WY.

Code information

UDI: 15099590799250/ V5.03 Panel Update 11

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2889-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2889-2026. Report a data issue: hello@regdatalab.com.