Z-2890-2020 Class II Ongoing

Recalled by Karl Storz Endoscopy — El Segundo, CA

Recall Details

Product Type
Devices
Report Date
September 2, 2020
Initiation Date
July 27, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
91

Product Description

Flexible Intubation Fiberscope, Part: 11301AA1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.

Reason for Recall

Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.

Code Information

Instruction Manual Version: 08/2018