Z-2938-2020 Class II Terminated

Recalled by Qiagen Sciences LLC — Germantown, MD

FDA device recall Z-2938-2020 was initiated by Qiagen Sciences LLC on Jul 15, 2020 and is designated Class II. Reason for recall: There is an increased rate of potential false positive results for the Influenza A (no subtype) target, which may lead to 1) inappropriate antiviral use, 2) erroneous or delayed differential diagnosis… The recall status is terminated (terminated Mar 22, 2021). Affected quantity: 1474 kits.

Recall details

Product Type
Devices
Report Date
Sep 16, 2020
Initiation Date
Jul 15, 2020
Termination Date
Mar 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1474 kits

Product description

QIAstat-Dx Respiratory SARS-CoV-2 Panel REF 691223 (US IVD under FDA EUA) - Product Usage: intended for the qualitative detection and differentiation of nucleic acid from multiple respiratory viral and bacterial organisms, including the SARS-CoV-2 virus, in nasopharyngeal swabs (NPS) eluted in universal transport media collected from patients suspected of COVID-19 by their healthcare provider.

Reason for recall

There is an increased rate of potential false positive results for the Influenza A (no subtype) target, which may lead to 1) inappropriate antiviral use, 2) erroneous or delayed differential diagnosis and missing of an alternative diagnosis.

Distribution pattern

US Nationwide distribution including in the states of AK, AL, CA, DC, DE, FL, GA, IL, IN, KY, MA, MD, MN, NC, NJ, NY, OH, OR, PA, TX, WA.

Code information

lots 166020729, 166022794, 166023368, 166023399, 166023400, 166024163, 166024165, 166024166, 166025078, 166025888, 166025889, 166028721, 166028722, 166029506, 166029508

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2938-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Qiagen Sciences LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-2938-2020. Report a data issue: hello@regdatalab.com.