Z-2973-2020 Class II Terminated

Recalled by Inpeco S.A. — Lugano, N/A

Recall Details

Product Type
Devices
Report Date
September 23, 2020
Initiation Date
August 7, 2020
Termination Date
December 2, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 US and 6 OUS

Product Description

FlexLab (FLX) with Sysmex XN-9000/XN-9100 Interface Module (dual robot) FLX-290-10; FLX-290-20 and Track to Rack Module FLX-295-10- Intended Laboratory Automation system - Product Usage: The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that perform a variety of instrument specific assays such as Sodium, Potassium and Chloride. Sodium, Potassium and Chloride measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.

Reason for Recall

Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of AK, FL, IA, LA, MA, MS, MT, NC, OH, TX, VA and the countries of AUSTRALIA, AUSTRIA, BELGIUM, DENMARK, FRANCE, GERMANY, ITALY, JAPAN, NETHERLANDS. NORWAY SAUDI ARABIA SOUTH KOREA SWEDEN UNITED ARAB EMIRATES

Code Information

Serial Numbers: US: FLX.0133 FLX.0181 OUS: FLX.0048 FLX.0049 FLX.0121 FLX.0155 FLX.0160 FLX.0200