Z-3012-2020 Class II Terminated

Recalled by Irvine Biomedical Inc, A St. Jude Medical Co. — Irvine, CA

FDA device recall Z-3012-2020 was initiated by Irvine Biomedical Inc, A St. Jude Medical Co. on Aug 24, 2020 and is designated Class II. Reason for recall: Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly… The recall status is terminated (terminated Jul 31, 2023). Affected quantity: 40 units.

Recall details

Product Type
Devices
Report Date
Sep 30, 2020
Initiation Date
Aug 24, 2020
Termination Date
Jul 31, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 units

Product description

Inquiry Steerable Diagnostic Catheter, REF: IBI-87002 - Product Usage: INDICATIONS FOR USE The Inquiry fixed curve and steerable electrophysiology catheters are used for electrogram recording and cardiac stimulation during diagnostic electrophysiology studies. The catheters are commonly placed at the high right atrium, right ventricular apex, and HIS bundle.

Reason for recall

Steerable diagnostic catheter packages labeled Model IBI-81104, batch 7397352 incorrectly contained Model IBI-87002, batch 7397307. Packages labeled Model IBI-87002, batch 7397307 incorrectly contained Model IBI-81104, batch 7397352.If physicians fail to observe the catheter difference, the intracardiac electrograms will not be representative of the desired signals, could result in procedure delay

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution including the states of MO, IN, WI, TX, AZ, MI, VA, KY and the country of Japan.

Code information

GTIN: 05414734309936, Lot: 7397307

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-3012-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Irvine Biomedical Inc, A St. Jude Medical Co.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Irvine Biomedical Inc, A St. Jude Medical Co.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Irvine Biomedical Inc, A St. Jude… →

Source: openFDA record for Z-3012-2020. Report a data issue: hello@regdatalab.com.