Z-3038-2018 Class II Terminated

Recalled by SPS Sterilization, Inc — Nixa, MO

Recall Details

Product Type
Devices
Report Date
September 19, 2018
Initiation Date
March 7, 2018
Termination Date
November 27, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units total

Product Description

Boston Scientific Wallstent Enteral Endoprosthesis Colonic/Duodenal Stent, 22MM X 90MM, resterilized.

Reason for Recall

The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.

Distribution Pattern

US distribution to AZ, CA, FL, OH, PA, TX, VA, WV

Code Information

Code Number 6559-R; Lot Number 17252887 (Serial Number 16211282016-02); Lot Number 17465176 (Serial Numbers 97403172017-16, 97403172017-40)