Z-3043-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 7, 2020
- Initiation Date
- August 18, 2020
- Termination Date
- September 30, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 170,425 units
Product Description
QuickTox Drug Screen DipCard(AMT Item/ AMT Description): QT70/ "QuickTox¿ Drug Screen Dipcard 11 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+MDMA500", QT88/ "QuickTox¿ Drug Screen Dipcard 12 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+ MDMA500+BUP10", QT11 / "QuickTox¿ Drug Screen Dipcard 5 Drug Detection: COC300+OPI2000+THC50+AMP1000+PCP25", QT52/ "QuickTox¿ Drug Screen Dipcard 9 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+MTD300+OXY100", QT44/ "QuickTox¿ Drug Screen Dipcard 8 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+OXY100", QT23/ "QuickTox¿ Drug Screen Dipcard 6 Drug Detection: COC300+OPI2000+MET500+THC50+AMP1000+PCP25", QT80F/ "QuickTox¿ Drug Screen Dipcard 12 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+TCA1000+OXY100+MDMA500 with Positive Control Line", QT52A/ "QuickTox¿ Drug Screen Dipcard 9 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+MTD300+OXY100 4 Adulteration Test: Cr, Ni, pH, Ox ", QT20/ "QuickTox" Drug Screen Dipcard 5 Drug Detection: COC300+OPI2000+MET500+THC50+AMP1000"
Reason for Recall
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Distribution Pattern
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Code Information
ALL Lots Within Expiry