FDA Inspection 1002222 — Apelem

Apelem — FEI 3000255482

The FDA completed an inspection of Apelem on January 26, 2017 in Gallargues-Le-Montueux. The inspection was classified VAI (Voluntary Action Indicated — objectionable conditions were found but do not meet the threshold for regulatory action). 7 citations were issued during this inspection. Product area: Devices.

Inspection Details

Classification
Voluntary Action Indicated (VAI) (VAI)
End Date
January 26, 2017
Fiscal Year
2017
Product Type
Devices
Project Area
Compliance: Devices
Location
Gallargues-Le-Montueux (France)

Citations

IDCFRDescription
1471221 CFR 820.184DHR - not or inadequately maintained
1471321 CFR 820.198(a)Lack of or inadequate complaint procedures
232721 CFR 820.22Quality audits - Lack of or inadequate procedures
310321 CFR 820.30(i)Design changes - Lack of or Inadequate Procedures
313021 CFR 820.100(a)Lack of or inadequate procedures
332321 CFR 820.170(b)Installer records
340921 CFR 820.200(d)Service reports